SCXI-1125 Calibration Procedure
Documentation
The NI-DAQ Function Reference Online Help file and NI-DAQ User
Manual for PC Compatibles 
contain detailed information on using the
NI-DAQ driver. The function reference online help file includes detailed
information on the driver functions. You can access the online help by
clicking Start»Programs»National Instruments DAQ»NI-DAQ Help.
The user manual provides instructions on installing and configuring
National Instruments DAQ devices. This manual also includes detailed
information on creating applications that use the NI-DAQ driver. These are
your primary references for writing your calibration utility. For further
information on the products you are calibrating, refer to your device user
manuals.
Test Conditions
Follow these guidelines to optimize the connections and the environment
during calibration:
Keep connections to the SCXI module short. Long cables and wires act
as antennae, picking up extra noise and thermal offsets that can affect
measurements.
Use shielded copper wire for all cable connections to the device.
Use twisted-pair wire to eliminate noise and thermal offsets.
Maintain the temperature between 18 and 28 °C.
Keep relative humidity below 80%.
Allow a warm-up time of at least 15 minutes for the SCXI module and
30 minutes for the E Series device to ensure the measurement circuitry
is at a stable operating temperature.
Calibration
The calibration process consists of the following steps:
1.
Setting up the device for testing.
2.
Verifying the existing operation of the device to determine whether it
is operating within its specifications.
3.
Adjusting the device with respect to a known voltage source.
4.
Verifying that the device is operating within its specifications after
adjustments.