SCXI-1125 Calibration Procedure
16. Record binary1 with volt1 and binary2 with volt2 for later use.
17. Repeat steps 7 through 16 for the remaining gain values you want to
adjust on this channel.
18. Repeat steps 6 through 17 for the remaining channels you want to
adjust.
You have determined the new calibration constants. Use these values in the
the following section,
.
Adjusting Calibration Constants
Complete the following steps to adjust the calibration constants on the
SCXI-1125:
1.
Call
SCXI_Cal_Constants
to create and store the new calibration
constants in the SCXI-1125 memory. Set the following parameters:
SCXIchassisID—The ID assigned by MAX
ModuleSlot—1, because the SCXI-1125 should be installed in
slot 1
channel—The channel on the SCXI-1125 you want to adjust
OpCode—2
calibrationArea—0
rangeCode—0 (not used for the SCXI-1125)
SCXIgain—The gain you want to adjust
DAQboard—The device number assigned by MAX
DAQChan—0
DAQGain—1 for 16-bit devices, –1 for 12-bit devices
Tbgain—1.0
volt1 and volt2—The first and second voltage readings for the
gain values that you are about to save on the SCXI-1125
binary1 and binary2—The first and second binary readings for
the gain values that you are about to save on the SCXI-1125
calConst1 and calConst2—These are return values
2.
Repeat step 1, changing calibrationArea to 1.
3.
Repeat steps 1 and 2 for the next channel.
You have now finished adjusting the calibration constants of your
SCXI-1125.