GE PT878GC 用户手册
Programming Site Data
4-53
April 2004
Entering Reynolds Correction (cont.)
e. When you have completed entering values, press
[F3]
(OK)
to confirm the table and return to the Correction Factors
window.
window.
You have completed entering data for a Reynolds Correction
table.
table.
•
To confirm the entries, press
[F3]
(OK). The meter returns to
Operate Mode.
•
To leave the window without confirming the entries and return
to Operate mode, press
to Operate mode, press
[F2]
(Cancel) or the
[ESC]
key.
Entering a Calibration Factor
The Calibration Factor is used as a reference for the flowmeter
electronics to ensure accurate readings. To enter a Calibration
Factor, return to the Reynolds Correction tab at the top of the
Correction Factors option. Press the [X] arrow key to reach the
Calibration Factor window. The window appears similar to
Figure 4-26 below. To step through each parameter, press the [T]
key.
electronics to ensure accurate readings. To enter a Calibration
Factor, return to the Reynolds Correction tab at the top of the
Correction Factors option. Press the [X] arrow key to reach the
Calibration Factor window. The window appears similar to
Figure 4-26 below. To step through each parameter, press the [T]
key.
Figure 4-26: The Calibration Factor Window
Reynolds Correction
Inputs
Correction Factors
Cancel
l
OK
Calibration Factor
Edit Table
Calibration Factor
On
Off
Meter K-Factor
Data Source
K Factor
Single
Table
10
Edit Table