Servicing the PT878GC
9-31
April 2004
Setting Up Pulse/Code Parameters (cont.)
2. The next prompt asks if you want to send a specific number of 
pulses. Press 
[ENTER]
 to select or deselect the pulse option.
Note: If you do not select this option, pressing the  [
T
] key 
returns the meter to the Transmit Code prompt.
3. The next prompt asks how many pulses (from 0 to 16) you 
wish to send.
a. Press 
[ENTER]
 to open the drop-down list.
b. Use the  [
T
] or [
S
] arrow keys to scroll to the 
recommended number.
c. Press 
[ENTER]
 to confirm your selection.
4. The final prompt, signal inversion, enables you to invert the  
transducer signal. Signal inversion may be necessary if your  
readings show unusual jitter, or for certain unusual 
combinations of pipe and fluid or pipe and lining. Signal 
inversion should be off, unless you have received other 
instructions. To change its status, use the [
W
 ] and [
X
] arrow 
keys to move to the appropriate radio button and press 
[ENTER]
.
You have completed entering data in the Signal Setup option.
Press 
[F2]
 (Cancel) to return to Operate Mode without 
confirming the changes.
Press 
[F3]
 (OK) to confirm the new data and return to 
Operate Mode.