Teledyne API M751H High Performance Zero Air Generator  
Operation
 
07773A DCN6833  
 
27 
4.2.17  Dewpoint Sensor  
The dewpoint sensor ensures that the Regenerative Dryer maintains a dewpoint less 
than -20°C (up to the maximum flow rate). This will increase the life of the chemical 
scrubbers.  When  the  dewpoint  drops  below  -20°C,  the  green  LED  (dewpoint 
indicator) on the instrument panel will light green, indicating that the air is dry. 
4.2.18  Instrument Panel Indicator Lights 
Multicolor indicator lights for both the power state and dewpoint state relay diagnostic 
information and instrument mode to the user.  
Table 4-1: Power and Dewpoint Condition Indicators
 
Power 
Indicator
 
Dew Point 
Indicator
 
Instrument State
 
Green
 
Green
 
Power ON, Dewpoint GOOD
 
Green
 
Yellow
 
Power ON, Dewpoint CAUTION 
 
Green
 
Red
 
Power ON, Dewpoint BAD
 
Green
 
Red-Flashing
 
Power ON, Dewpoint CRITICAL
 
Alternating 
Yellow – 
Green 
Representative 
of Dewpoint 
State 
Maintenance Mode.  24hr Dewpoint Fault 
Override 
Red-Flashing
 
Yellow
 
Temperature Fault.  
 
Red-Flashing
 
Red
 
Dew Point Fault.