ResMed VPAP Adapt SV 用户手册

下载
页码 42
33
Technical Specifications
English
SmartStop
A feature that stops the VPAP Adapt SV automatically when you remove the 
mask. You can turn SmartStop on or off in the Settings menu.
Tidal volume
The volume of air, in liters, breathed in or out in one breath.
Technical Specifications
Dimensions (H x W x D)
5.6" x 9.8" x 11.5" (142 mm x 248 mm x 293 mm)
Weight
Flow generator with cover: 8.14 lbs (3.7 kg)
Flow generator with HumidAire 2i (empty): 9.24 lbs (4.2 kg)
Environmental conditions
Operating temperature: 41
º
F to 95
º
F (+5
º
C to +35
º
C)
Storage and transport temperature: -4
º
F to +140
º
F (-20
º
C to +60
º
C)
Humidity—operating, storage and transportation: 10%–95% non-condensing
Housing construction
Injection moulded plastic
Air tubing
1 x 6’6” (2 m)
Air filter 
Synthetic fiber
Patient connection port
22 mm taper, compatible with ISO 5356-1:2004 Anaesthetic & Respiratory 
Equipment—Conical Connectors
Power supply
AC Input range: 110–120V and 220–240V; 50–60Hz; 60VA 
DC Input range (if using DC Converter: 12–13.8V DC
Performance
Operating pressure range: 4–25 cm H
2
O
Sound pressure level
<30 dBA (tested in accordance with the requirements of ISO 17510-1)
Battery type
Lithium
228234 VPAP Adapt SV User Guide USA Eng.book  Page 33  Monday, October 13, 2008  2:15 PM