ResMed VPAP Adapt SV 用户手册

下载
页码 42
36
Guidance and Manufacturer’s Declaration — Electromagnetic 
Emissions and Immunity 
Guidance and manufacturer’s declaration – electromagnetic emissions
The VPAP Adapt SV is intended for use in the electromagnetic environment specified below. The 
customer or the user of the VPAP Adapt SV should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - 
guidance
RF emissions CISPR 11 
Group 1
The VPAP Adapt SV uses RF energy 
only for its internal function. 
Therefore, its RF emissions are very 
low and are not likely to cause any 
interference in nearby electronic 
equipment.
RF emissions CISPR 11 
Class B
The VPAP Adapt SV is suitable for use 
in all establishments, including 
domestic establishments and those 
directly connected to the public low-
voltage network that supplies 
buildings used for domestic purposes.
Harmonic Emissions 
IEC 61000-3-2
Class A
Voltage Fluctuations/Flicker 
Emissions IEC 61000-3-3
Complies
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed 
and put into service according to EMC information provided in this document.
Warnings: The VPAP Adapt SV should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the VPAP Adapt SV should be observed to verify normal 
operation in the configuration in which it will be used.
The use of accessories (eg Humidifiers) other than those specified in this manual is not 
recommended. They may result in increased emissions or immunity of the VPAP Adapt SV.