ResMed VPAP Adapt SV 用户手册

下载
页码 42
4
impaired ability to clear secretions
pneumothorax or pneumomediastinum
recent cranial trauma or surgery
chronic hypoventilation.
Below are general warnings and cautions. Further specific warnings, 
cautions and notes appear next to the relevant instructions in the manual.
!
WARNINGS
The VPAP Adapt SV is NOT a life support ventilator.
The entire manual should be read before using the VPAP Adapt SV.
Advice contained in this manual should not supersede instructions given by 
the prescribing physician.
The VPAP Adapt SV should be used with masks and accessories 
recommended by ResMed or the prescribing physician. Use of incorrect 
masks and accessories may adversely affect the function of the 
VPAP Adapt SV. 
The VPAP Adapt SV is designed for use with masks that allow exhaled gases 
to be flushed out through vent holes. Exhaled gases will be rebreathed if the 
mask is worn with the machine turned off, or the vent holes are occluded. If 
this occurs over prolonged periods, suffocation may occur.
In the event of power failure or machine malfunction, remove the mask.
The air flow for breathing produced by this device can be as much as 11
º
(6
º
C ) higher than the temperature of the room. Caution should be exercised 
if the room temperature is warmer than 90
º
F (32
º
C).
The VPAP Adapt SV can be set to deliver pressures up to 25 cm H
2
O. In the 
unlikely event of certain fault conditions, pressures of up to 40 cm H
2
O for 
less than one second are possible.
The VPAP Adapt SV is not suitable for use in the vicinity of flammable 
anesthetics.
If oxygen is used with the VPAP Adapt SV, the oxygen flow should be 
stopped when the device is not operating. If oxygen flow continues when 
the device is not operating, oxygen may accumulate within the device and 
create a risk of fire.
Do not use the VPAP Adapt SV if there are obvious external defects or 
unexplained changes in performance.
Do not open the VPAP Adapt SV case. There are no user serviceable parts 
inside. Repairs and internal servicing should only be performed by an 
authorized service agent.
!
CAUTIONS
You should report unusual chest pain, severe headache or increased 
breathlessness to your physician. An acute upper respiratory tract infection may 
require temporary discontinuation of treatment.
228234 VPAP Adapt SV User Guide USA Eng.book  Page 4  Monday, October 13, 2008  2:15 PM