NI Digital Waveform Generator/Analyzer Guide
Tables 8 and 9 describe the different conditions indicated by the two 
NI PXI LEDs.
1, 4, 7, 10, 13, 
16, 19, 22, 25, 
28, 31, 34, 37, 
40, 43, 46, 49, 
52, 55, 58
GND
Ground
Ground reference for signals.
11, 12, 68, 69
RESERVED
N/A
These terminals are reserved for 
future use. Do not connect to 
these pins.
Table 7.  NI 654X/655X/656X SMB Connectors
Connector
Signal 
Name
Signal 
Type
Description
CLK IN
Reference/
Clock Input
Control
Terminal for the external Reference 
clock used for the PLL or for the 
external Sample clock used for 
pattern generation and/or 
acquisition.
PFI 0
Programmable 
Function 
Interface 
(PFI) 0
Control
Input terminal to the device for 
external triggers, or output terminal 
from the device for events.
CLK OUT
Reference/
Clock Output
Control
Terminal for the exported PLL 
Reference clock or the exported 
Sample clock.
Table 8.  ACTIVE LED Indicators
Color
Indications
Off
Device not armed, not triggered, or experiencing an error.
Amber
Device armed and awaiting Start trigger. If performing a 
dynamic acquisition operation, the device may be acquiring 
pretrigger samples.
Green
Device received Start trigger.
Red
Error condition.
Table 6.  NI 656X DDC Connector Pins (Continued)