NI Digital Waveform Generator/Analyzer Guide
Device Does Not Appear in MAX
Complete the following steps if the NI device does not appear in MAX:
1. In the MAX Configuration pane, click Devices and Interfaces to expand 
the category.
2. Click NI-DAQmx Devices and press <F5> to refresh the list of installed 
devices.
3. If the device is still not listed, power down the system, ensure the device 
is correctly installed, and restart.
4. If the device still does not appear under NI-DAQmx Devices, contact 
NI support at 
ni.com/support
.
Device Failed the Self-Test
The MAX self-test performs a brief test of device resources. If the device 
does not pass the self-test, complete the following steps:
1. Reboot your system.
2. Launch MAX and perform the self-test again. If the device still fails the 
self-test, proceed to step 3.
3. Uninstall and reinstall NI-HSDIO.
4. If the device still fails the self-test, contact NI support at 
ni.com/
support
.
I Received a Thermal Shutdown Error
I received the 
kErrorDeviceShutDownDueToHighTemp
 error, and my 
device shut down. What should I do next?
To re-enable your device after a thermal shutdown, you must perform a hard 
reset, in which the device integrated circuits (ICs) are reloaded. To re-enable 
your device after thermal shutdown, complete the following steps:
1. Power down the computer or chassis that contains the device.
2. Review the procedure in step 
 and make any 
necessary adjustments to ensure that your device is effectively cooled.
3. Call 
niHSDIO_ResetDevice
 (NI-HSDIO Reset Device VI) or perform 
a device reset in MAX. For more information on performing a device 
reset in MAX, refer to step
The thermal shutdown error continues to be reported until the device has 
been successfully reset.