National Instruments NI PXI-5422 用户手册

下载
页码 63
© National Instruments Corporation
31
NI PXI-5422 Calibration Procedure
4.
Call 
niFgen_SetAttributeViReal64
 to set the gain (niFgen 
property node: Arbitrary Waveform Output»Arbitrary 
Waveform Gain
) using the following parameters:
vi: The session handle returned from 
niFgen_init
channelName
"0"
attributeID
NIFGEN_ATTR_ARB_GAIN
value: The Gain value listed in Table 6 for the current iteration
5.
Call 
niFgen_SetAttributeViInt32
 to choose the positive 
full-scale DC waveform (niFgen property node: Arbitrary 
Waveform Output»Arbitrary Waveform Handle
) using 
the following parameters:
vi: The session handle returned from 
niFgen_init
channelName
"0"
attributeID
NIFGEN_ATTR_ARB_WAVEFORM_HANDLE
value: The wfmHandle from step 9 of th
 section (positive full-scale handle)
6.
Call 
niFgen_InitiateGeneration
 (niFgen Initiate Generation VI) 
to initiate the waveform generation using the following parameter:
vi: The session handle returned from 
niFgen_init
 
7.
Measure the positive DC voltage out of the NI PXI-5422. Record the 
positive DC voltage out measurement in the Measured Positive 
Full-Scale Value 
column of Table 6. 
8.
Call 
niFgen_AbortGeneration
 (niFgen Abort Generation VI) 
to abort the waveform generation using the following parameter:
vi: The session handle returned from 
niFgen_init
9.
Call 
niFgen_SetAttributeViInt32
 to choose the negative 
full-scale DC waveform (niFgen property node: Arbitrary 
Waveform Output»Arbitrary Waveform Handle
) using 
the following parameters:
vi: The session handle returned from 
niFgen_init
channelName
"0"
attributeID
NIFGEN_ATTR_ARB_WAVEFORM_HANDLE
value: The wfmHandle from step 11 of the 
 section (negative full-scale handle)
10. Call 
niFgen_InitiateGeneration
 (niFgen Initiate Generation VI) 
to initiate the waveform generation using the following parameter:
vi: The session handle returned from 
niFgen_init