National Instruments NI PXI-5422 用户手册

下载
页码 63
© National Instruments Corporation
35
NI PXI-5422 Calibration Procedure
Initializing the External Calibration Session
Call 
niFgen_InitExtCal
 (niFgen Init Ext Cal VI) to open an NI-FGEN 
external calibration session using the following parameters:
resourceName: The name of the device you want to calibrate. 
This name can be found under Devices and Interfaces in MAX.
password: The password required to open an external calibration 
session. If this password has not been changed since manufacturing, 
the password is 
NI
.
vi: A pointer to a ViSession. The variable passed by reference through 
this parameter receives the value that identifies the external calibration 
session created by this function. This value acts as the session handle 
and is passed as the first parameter to all subsequent NI-FGEN 
functions.
Adjusting the Analog Output
The analog output adjustment procedure has several sub-procedures that 
adjust the following:
Main path pre-amplifier offset
Main path pre-amplifier gain
Main path post-amplifier gain and offset
Direct path gain
In each of these sub-procedures, you put the device in several 
configurations and take several output measurements. You then pass 
these measurements to NI-FGEN, which determines the calibration 
constants for the device.
Initializing Analog Output Calibration
1.
Call 
niFgen_InitializeAnalogOutputCalibration
 (niFgen 
Initialize Analog Output Calibration VI) using the following 
parameter:
vi: The session handle returned from 
niFgen_InitExtCal
 
2.
Call 
niFgen_WriteBinary16AnalogStaticValue
 (niFgen Write 
Binary 16 Analog Static Value VI) to set the main DAC value using 
the following parameters:
vi: The session handle returned from 
niFgen_InitExtCal
 
channelName
"0"
value
0