NI PXI-5422 Calibration Procedure
6
ni.com
Documentation
For information about NI-FGEN and the NI PXI-5422, you may find the 
following documents helpful:
NI Signal Generators Getting Started Guide
NI Signal Generators Help
NI PXI-5422 Specifications
NI-FGEN Instrument Driver Quick Reference Guide
The NI Signal Generators Getting Started Guide provides instructions for 
installing and configuring NI signal generators. 
The NI Signal Generators Help includes detailed information about 
the NI PXI-5422 and the NI-FGEN functions. You can access this help file 
by selecting Start»Programs»National Instruments»NI-FGEN» 
Documentation»NI Signal Generators Help
. For the latest versions 
of NI documentation, refer to 
ni.com/manuals
.
Self-Calibration Procedures
The NI PXI-5422 is capable of performing self-calibration, which adjusts 
the gain and offset of the main and direct analog paths. Self-calibration 
exclusively uses an onboard A/D converter (ADC) to measure the output 
voltage. You can implement self-calibration on the NI PXI-5422 by 
following procedures similar to th
 and the 
 procedures. However, 
output impedance, oscillator frequency, and the calibration ADC cannot be 
adjusted during self-calibration.
You can initiate self-calibration interactively from Measurement & 
Automation Explorer (MAX) or from the FGEN Soft Front Panel (SFP). 
Alternately, you can initiate self-calibration programmatically using 
NI-FGEN. 
MAX
To initiate self-calibration from MAX, complete the following steps:
1.
Launch MAX.
2.
Select My System»Devices and Interfaces»NI-DAQmx Devices.