Tektronix TDS460A 用户手册
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–257
Save and Recall
DELEte:SETUp, RECAll:SETUp, *RCL, *SAV
SAVe:SETUp { <NR1> | <file path> }
<NR1>
SAVe
:
SETUp
<Space>
<file path>
<NR1> is a value in the range from 1 to 10 and specifies a location. Using an
out-of-range value causes an execution error. Any settings that have been stored
previously at this location will be overwritten.
out-of-range value causes an execution error. Any settings that have been stored
previously at this location will be overwritten.
<file path> (available on instruments with the Option 1F File System) is a
quoted string that defines the file name and path. Input the file path using the
form
quoted string that defines the file name and path. Input the file path using the
form
<drive>/<dir>/<filename>. <drive> and one or more <dir>s are
optional. If you do not specify them, the TDS will write the file to the current
directory.
directory.
<filename> stands for a filename of up to 8 characters followed by a
period (“.”) and a 3-char extension . We recommend you use “
SET” for the
extension to identify files that store setup data.
The current directory refers to the name of a directory as returned by the
FILESystem:CWD query.
FILESystem:CWD query.
Settings saved in one TDS oscilloscope may or may not work on a different
model TDS or on the same model TDS with a different version of firmware.
model TDS or on the same model TDS with a different version of firmware.
SAVE:SETUP 5
saves the current front-panel setup in memory location 5.
saves the current front-panel setup in memory location 5.
SAVE:SETUP TEK00000.SET
saves the current front-panel setup to the file
saves the current front-panel setup to the file
TEK00000.SET in the current
directory and on the current drive.
SAVe:WAVEform (No Query Form)
Stores a waveform in one of four reference memory locations or a mass storage
file. This command is equivalent to selecting either the Save Waveform or the
Save to File Waveform item in the Save/Recall Waveform menu.
file. This command is equivalent to selecting either the Save Waveform or the
Save to File Waveform item in the Save/Recall Waveform menu.
Save and Recall
Group
Related Commands
Syntax
Arguments
Examples
Group