oxyCRG Display
3-26     Setting up your Monitor  
S
e
tt
ing up y
our
 
M
onito
r  
Approximately the bottom two thirds of the Main Screen area are 
covered with the three oxyCRG waveforms.
The number of realtime waves that you can display simultaneously with 
oxyCRG is related to the maximum number of realtime waves that your 
system is capable of displaying. Details are provided in the following 
table:
Main Screen
Notes on 
oxyCRG
The oxyCRG information is overlayed on the realtime waves at the 
bottom of the Main Screen. These realtime waves disappear from 
the display, but the parameters are still being monitored, and the 
parameter alarms are still active.
If oxyCRG covers part of an overlap channel, the entire overlap 
channel disappears from the display.
The 
 
softkey has no effect on the oxyCRG section 
of the Main Screen, because the display speed is very low.
The beat-to-beat heart rate (btbHR) in the oxyCRG display will 
vary from the heart rate (HR) numeric on the Main Screen, since 
the HR numeric is an averaged value.
If the monitor is switched off for less than 60 seconds, oxyCRG 
data is retained. However, when the power is switched back on, 
you will see gaps in the oxyCRG display. The gaps correspond to 
the amount of time it takes the monitor to get the first readings 
after power is restored, not the amount of time that the monitor 
was switched off. For this reason, you cannot properly measure 
the time difference between an event which occurs before power 
Standard Display
oxyCRG Display
4 realtime waves
oxyCRG + 1 realtime wave
6 realtime waves
oxyCRG + 2 realtime waves
8 realtime waves
oxyCRG + 3 realtime waves
Freeze Waves