Philips V24CT 用户手册

下载
页码 392
Configuring Module, Bedside and Central Recordings
3-40     Setting up your Monitor  
S
e
tt
ing up y
our
 
M
onito
r  
Configuring Module, Bedside and Central Recordings
Note—
Recordings on a 4-channel bedside recorder are only possible 
with the CMS.
The 
 key enables you to configure five types of 
recordings:
1. Delayed and Alarm Recordings on a module, bedside or central 
recorder. A maximum of four delayed waves can be recorded, 
depending on the type of recorder. Up to four additional waves 
can be selected for alarm recordings, depending on the model of 
the monitor.
2. Realtime Recordings. Three recording modes can be configured to 
record waveforms at your request. For each mode you can choose 
the waves to be recorded, their position on the recording strip and 
the recorder to be used. Realtime recordings can be produced on 
module, 4-channel bedside or central recorders.
Note—
To ensure an optimal realtime recording check that the 
recording configuration is appropriate for your recorder. 
3. Monitoring Procedure Recordings on a module, bedside or central 
recorder. You can record the curves produced during cardiac 
output and pulmonary artery wedge pressure measurements.
4. Vital Signs Recordings. Automatic Vital Signs Recording gives the 
option of printing out all parameter numerics on the Plug-in 
Recorder Module either at user programmed time intervals, or 
upon completion of an NBP measurement.
5. OxyCRG and Neonatal Event Review Recordings.You can record 
the contents of the oxyCRG display and Neonatal Events onto an 
M1116B plug-in recorder module. For Neonatal Event Review 
Recordings, the M1116B plug-in recorder module has to be 
selected as the procedure recorder.
The procedures for configuring recordings are described in Chapter 6 
“Recording Functions”
.
Recordng Setup