Philips V24CT 用户手册

下载
页码 392
Changing the Patient Category
3-48     Setting up your Monitor  
S
e
tt
ing up y
our
 
M
onito
r  
These safety values effect the measurement range of each category. The 
Adult category has the greatest range (systolic 30 to 270 mmHg) and the 
Neonate has the smallest (systolic 30 to 130 mmHg). Therefore you 
cannot use Neonate or Pediatric if the expected pressure values 
are outside the measurement range for that category.
 The following tables list the measurement ranges and safety values for 
each patient category.
NBP Measurement Range (in mmHg) for Differing Patient 
Categories
Measurement Range
Patient Category
Adult
Pedi
Neo
Systolic
30 to 270
30 to 180
30 to 130
Mean
20 to 255
 20 to 160
20 to 120
Diastolic
10 to 245
10 to 150
10 to 100
Cuff Size 
A to E 
Any Cuff 
1 to 4 & A
Default Inflation Pressure
165 mmHg 
125 mmHg
100 mmHg
NBP Safety Values for Differing Patient Categories
Safety Value
Patient Category
Adult
Pedi
Neo
Continuous Pressure
15 mmHg 
(for 120s)
5 mmHg 
(for 120s)
5 mmHg (for 
60s)
Time to inflate 
from 0 to 60 mm Hg
 40s 
 20s 
 20s
Max. inflation time 
60s   
40s
30s