Philips V24CT 用户手册

下载
页码 392
Parameter Settings Transfer Messages
Setting up your Monitor   3-63
S
e
tt
ing up y
our
 
M
onit
o
r  
Parameter Settings Transfer Messages 
The following messages related to Parameter Settings Transfer can 
appear on the display:
Message 
 Condition 
 Action required
(Parameter name) 
no settings 
transfer capability
The indicated parameter 
module is not a T 
module. 
Insert a module that has 
T on the front.
(Parameter name) 
check settings'' 
No settings are stored in 
the module, or the 
settings are incorrectly 
stored. 
Enter the parameter 
Task Window and check 
the settings. The   
settings are 
automatically stored 
when you exit the Task 
Window.
Check Patient 
Category and 
(parameter name) 
settings
The patient category 
stored with the settings 
in the module is 
different from that of the 
monitor. 
Either
 change the 
patient category for the 
monitor or change   the 
parameter settings.
(Parameter name) 
unable to store 
settings in module
The parameter module 
was removed before the 
settings could be stored, 
or there is a malfunction 
in the module. (This 
message can appear with 
Settings Transfer On or 
Off.) 
Press , 
 
 
re-insert the module and 
check/adjust the 
settings in the   
parameter Task 
Window. Exit the Task 
Window and leave the 
module   in the rack for 
about 10 seconds before 
you remove it.   If the 
same message appears, 
replace the module.
Silence Reset