Philips Patient Care System
Other Patients   4-7
Ot
h
e
r Pa
ti
e
n
ts
  
Configuring 
the Other 
Patients 
Controls
With the Other Patients Controls you can place your own bed into one of 
eight groups: A, B, C, D, E, F, G or the ALL group, based on nursing care 
assignments within a unit.
The Other Patients Controls Task Window allows you to change the Send/
Receive alarms configuration for your bed.
Note—
When a bed is assigned to the ALL group, alarms can be received 
from and sent to beds in group A-G as well as the other beds in the ALL 
group. When a bed is assigned to any group in A-G, alarms are received 
from and sent to any bed in the same group and also to beds in the ALL 
group. 
Note—
The bed will automatically be assigned to the ALL group if the 
connection to the Philips monitoring network is interrupted (for example, 
if the Philips monitoring network cable at the back of the monitor is 
disconnected, or the central station is switched off) and the monitor 
comes back on a different branch.
Note—
Any changes made to Other Patients settings will be stored for up 
to three hours after the monitor is switched off.