Alarm Display
5-4     Alarm Functions  
Ala
rm Func
tions
  
If all the alarms have been turned off using the 
 
softkey or the 
 hardkey, they can all be turned back on by 
pressing the 
 softkey or the 
 hardkey.
While the alarms are suspended (permanently or for 1, 2 or 3 minutes), 
INOP messages continue to be displayed but alarm messages, lamps and 
alarm tones are not active. The nurse call relay is inactive.
Suspended
Alarms during
Arrhythmia
Monitoring
Notes—
While Arrhythmia is assigned, the central Arrhythmia Computer 
takes over the ECG/HR alarms.
 Suspending Alarms switches off all alarms, including Arrhythmia 
alarms.
If Arrhythmia is assigned and alarms are suspended, the ECG/HR 
alarms are switched off. The 
 Softkey in the Alarm Limits 
and ECG Adjust Alarms Task Windows will not be available 
(inactive softkey). The ECG/HR alarm 
 capability can 
only be reactivated by switching the main Alarms on, or 
unassigning Arrhythmia. 
Silencing 
and 
Resetting 
Alarms
If an alarm occurs you can silence it by pressing the 
 
key. This will silence ALL alarms that are occurring at that moment. The 
alarm messages numeric will continue to blink.
You can configure your system on installation to have all latching or all 
non-latching alarms or to have visual alarms latching and audible alarms 
non-latching. You can also configure the alarm reminder function to be 
on or off.
Note—
We recommend that you configure latching alarms for your 
monitor for Intensive Care use. All monitors within a unit should have 
the same alarm configuration!
Note—
Audible alarm indications broadcast over the Philips monitoring 
network (SDN) are always treated like visual alarms on the central 
station or remote systems (even if the bedside monitor is configured 
differently).
Suspend Alarms
Suspend
 
SwitchOn Alarms
Suspend
 
ON/OFF
ON/OFF
Silence/Reset