Alarm Display
5-8     Alarm Functions  
Ala
rm Func
tions
  
Alarm 
Priorities
Red alarms -
 Highest priority alarms, displayed in the top right 
corner of the display screen. These are marked with three 
asterisks (***) next to the alarm message, for example, 
***ASYSTOLE.
Yellow alarms -
 displayed in the top center of the display screen. 
There are two types of yellow alarms; continuous yellow alarms 
and short yellow alarms. The continuous yellow alarms are 
associated with preset alarm limits. The short yellow alarms are 
active from the arrhythmia computer.
Both types of yellow alarms are marked with two asterisks (**) 
next to the alarm message, for example, **HR 56<60, but the alarm 
tone is shorter with the alarms from the arrhythmia computer. For 
both the short and continuous yellow alarm, the yellow lamp is lit.
INOPs - 
displayed in the top left corner of the display screen. The 
INOPs also have priority levels, they are as follows:
Equipment Related INOPs - These are caused by setup or 
hardware faults, for example, “ABP NO TRANSDUCER”. You 
will hear an INOP tone. When an equipment-related INOP can 
have a direct effect on the patient, for example “NBP CUFF 
OVERPRESS” of the noninvasive pressure cuff, it is 
categorized as a severe| INOP. When a severe INOP occurs 
the system alarm capability is switched ON by the system, if 
not already on. You will hear an INOP tone. System alarms 
cannot be suspended until the cause of the severe INOP has 
been eliminated.
Patient Related INOPs - These are usually caused by the 
system being unable to process signals properly, for example, 
“Noisy ECG”. You will not hear an INOP tone with this level 
of INOP.