Alarm Setup
Alarm Functions   5-15
Ala
rm Func
tions
  
The Nurse 
Call Relay
Note—
The nurse call relay is not available for the V24CT and V26CT.
The nurse call relay is a paging system which warns the nurse if an alarm 
occurs on a monitor. It becomes active in one of four situations 
depending on the way it has been configured:
Red only alarms
Red and Yellow alarms
Red and INOP alarms
Red and Yellow INOP alarms.
The nurse call relay is not active when the monitor’s alarm capability is 
switched off.
Note—
The delay time to the nurse call relay output is max. 1 sec.
When an alarm has been silenced using the 
 key, and the 
audible 2 or 3 minute alarm reminder is configured to be either On or Re-
alarm, the nurse call relay signals will behave as follows:
Alarm 
Reminder
Reminder 
Time
Nurse Call Relay Signal Behavior
On
or 3 mins
The nurse call relay signals are repeated 
for a duration of 6 seconds every 2 or 3 
minutes unless the alarm condition is 
eliminated
Re-alarm
or 3 mins
The nurse call relay signals are activated 
after 2 or 3 minutes. The alarm tone and 
relay signals run continuously until they 
are interrupted by pressing the 
 key.
Silence/Reset
 
Silence/Reset