Alarm Recording
6-14     Recording Functions  
Re
c
o
rd
ing
 Func
tions
  
Alarm 
Recording 
Priorities
In a delayed recording, waves are recorded in the order assigned in the 
Delayed/Alarm Recording Task Window, and the number of waves 
recorded is determined by the recorder configured for delayed 
recordings. In alarm recordings, the alarming parameters replace the 
waves in the lower channels. In the default alarm recording, the top 
channel contains the first delayed wave, usually ECG. Additional 
delayed channels are overwritten by the additional alarming parameter, 
starting at the bottom channel, in the following priority:
Parameter whose alarm status has escalated in the last second, 
and has alarm recording enabled. 
Parameter in red alarm with alarm recording enabled.
Parameter in yellow alarm with alarm recording enabled. 
Parameter in red alarm with alarm recording disabled (if space is 
available). 
Parameter in yellow alarm with alarm recording disabled (if space 
is available).
Alarm
Recording
Examples
On the Plug-In (2-channel) recorder with respiration recordings 
disabled, an apnea alarm followed by a Pressure 1 limit alarm would 
produce a strip:
provided that P1 has been selected in the Delayed/Alarm Recordings 
Task Window.
For a 4-channel bedside recorder with a standard delayed recording 
configured as:
Channel 1 
 ECG 
Channel 2 
Pressure 1 
Channel 1
ECG
Channel 2
Pressure 1
Channel 3
Pressure 2