Additional Information
6-40     Recording Functions  
Re
c
o
rd
ing
 Func
tions
  
For example, the complete code (8AMIN360) at the beginning of a 
recording strip means: 
The recording is delayed (8A). 
The monitor is in monitoring mode (M). 
The application area is ICU (I). 
The patient type is neonatal (N). 
The delay time is 6 minutes, and the recording is an oxyCRG one 
(360). 
Realtime and delayed recordings are also annotated with the following 
information:
Date and time (time of the alarm for alarm recordings, time of 
request for delayed recordings, time data was measured for stored 
recordings such as cardiac output, and time when run for others). 
Alarm/INOP text (if applicable) 
Patient  parameters 
Recorder speed 
Waveform labels 
Scale lines and labels 
Bed label and patient name (if the monitor is attached to the 
Philips patient care system, or if the patient data management 
option has been purchased). 
Note—
For recordings made at speeds lower than 6.25 mm/s, the time is 
updated every 15 minutes. If you are making a continuous realtime 
recording at these speeds on the 4-channel recorder, the recording is 
started at 25 mm/s for the calibration marks and annotations, then 
switched to the configured speed. No re-annotations are made. 
Recordings made at speeds greater or equal to 6.25 mm/s are annotated 
every 60 seconds on the Plug-In and 4-channel recorders. There is no re-
annotation on central station recorders. 
A calibration pulse scale is included for ECG, pressure, CO
2
, EEG and 
VueLink waveforms. This annotation appears above the waveform. For 
cardiac output recordings, the calibration annotation takes the form: