Recording Functions   6-43
Re
c
o
rd
ing
 Func
tions
  
Changing 
the 
Recorder 
Speed
The recording speed can be changed for delayed, realtime, and oxyCRG 
recordings made on a bedside recorder. Changes made in the Change 
Recording Speed window affect only the last requested and running 
recording.
Note—
If more than one bedside recording is running, the Change 
Recording Speed window affects only the last requested and running 
recording. However, if the last requested recording is stopped, the Change 
Recording Speed window then affects the recording that started running 
before the stopped one. 
Because the recorder must complete the annotation at the beginning of 
the strip before the key to access this window appears, there may be a 
wait of a few seconds before the speed can be changed. For delayed 
recordings, access to the Change Recording Speed window is through the 
 key. The new speed is not stored.
The speed of alarm recordings cannot be changed; these recordings 
always run at 25 mm/sec.
Changing 
the 
Recorder
You can change the recorder where alarm, delayed, preset, or realtime 
non-configured recordings are made.
1. Press . 
2. Press . 
3. Select the recording type (Delayed & Alarm, Mode, or Procedure). 
4. Press 
 to choose the appropriate recorder. 
Continuing 
a Timed 
Recording 
Press 
on the recorder. The recording runs until you stop it (by 
pressing the 
 on the recorder or the 
 softkey on 
the display).
Realtime Record
Monitor Setup
Recordng Setup
Change Recorder
CONT
 
STOP
Stop Recordng