Philips V24CT 用户手册

下载
页码 392
Viewing Patient Data
8-10     Trends and Calculations  
Tr
e
nds
 a
nd 
Calcula
tion
s
  
For example, when you press 
 from a Vital Signs display 
at one-minute resolution, the total time the graphs spans is 1 hour. 
Conversely, when you go directly from Graph Trends with a 2-hour span 
to the Vital Signs display, the intervals are 15-minute. 
Note—
If configured for the OR environment, the data is stored every 12 
seconds. Depending on the patient data management configuration and 
the options purchased, the default data time intervals are 1-min., 5-min., 
15-min., and 1-hr. with a maximum of 4 hours; and (CMS only) 1-min., 5-
min., 15-min., 1-hr., and 2-hrs. (with a maximum of 9 hours). 
Up to 13 lines are displayed on the screen at one time. Only monitored 
parameters are shown. The continuously monitored parameters are 
shown in the general order in which their numeric values appear on the 
Main Screen, followed by the aperiodic parameters (for example, C.O.). 
Triple-valued parameters, such as ABP, are shown on three separate 
lines. Parameter units (for example, beats/minute) are not shown. 
Note—
If the numeric for a parameter is turned OFF in the Display Setup 
/ Numerics On/ Off Task Window (but the parameter is ON), the data is 
still stored and will be shown in the table. 
Graph Trends