Viewing Patient Data
8-18     Trends and Calculations  
Tr
e
nds
 a
nd 
Calcula
tion
s
  
Procedure to Insert a Calibration Pulse
 
 
 
 
When you press 
, you'll hear a beep as the calibration 
pulse appears in the displayed waves. The event label softkeys and the 
 softkey disappear. In the time period between the 
 softkey being pressed and it disappearing from the 
screen, parameter numeric values are not collected and alarms are not 
active. The calibration pulse will stay on the screen for the length of 
time it takes a waveform to move across the screen. If the waveform is 
fixed, the calibration pulse will be overwritten. The 
 
key 
under
 
 can be used to freeze the screen to allow close 
examination of the waveform(s) and calibration signals.
Pressing 
 during a recording produces a calibration pulse 
on that recording (in addition to the calibration wave that routinely 
appears at the beginning of the recording). 
The calibration pulse does not appear in the Vital Signs or Graphs 
display.
Trends/Calcs
Mark Event
Cal Signals
Cal Signals
Cal Signals
Cal Signals
Freeze Waves
Change Screen
Cal Signals