Performing and Reviewing Calculations
8-24     Trends and Calculations  
Tr
e
nds
 a
nd 
Calcula
tion
s
  
Resampling
Vital SIgns
Resampling of vital signs enables you to override the values in the 
Trends database, which are stored every 12 seconds or averaged over 1 
minute, depending on the monitor’s configuration. Resampling enables 
you to use the most recent continuously monitored values. Resampling 
sets the calculation time to the current time, and displays the 
corresponding values for the previous second in the calculations task 
window. Resampled values are indicated by a preceding asterisk (*) in 
Calculations, Graph Trends, and Vital Signs task windows.
If no current values exist, then the most recent value within 30 minutes 
will be retrieved. The most recent height, weight, (and head 
circumference for neonatal only) are used regardless of how long ago 
they were entered.
Pressing 
 stores the resampled values at the current time. 
Pressing 
 
also stores the resampled values at the current 
time and does the calculations using the resampled values. Pressing any 
other softkey except 
 
cancels the resampled values and 
brings up the database values for the displayed calculation time. 
Resampled values that are not stored or used in calculations will be 
erased.
When data is transferred with the Data Transfer module, values that 
have been resampled are transferred to the destination monitor. 
Resampled and other manually entered data is annotated with an 
asterisk (*) at the destination monitor if both the source and destination 
monitors have software at the Release F level or higher (CMS) or 
Release A.0 level or higher (V24 and V26).
Confirm
Perform Calc
On/Off DataTime