Philips V24CT 用户手册

下载
页码 392
Viewing oxyCRG Episodes
9-16     Neonatal Event Review  
Ne
ona
ta
l Ev
e
n
Revi
ew
  
The exact time an event was triggered is indicated by a vertical line 
which is positioned on the graph according to the monitor's 
configuration. Depending upon the user's requirements, the monitor can 
be configured (in Configuration Mode) to display and store an oxyCRG 
Episode in the following way:
Two minutes of Pre-Event Time / Two minutes of Post-Event Time 
(the default) or 
One minute of Pre-Event Time / Three minutes of Post-Event Time
Note—
In both cases manual events are displayed with four minutes Pre-
Event Time.
The association of events is determined by the configuration of the post 
event time. The events that occur during the configured two or three 
minutes of post event time are associated and displayed as one event in 
the Event Review and oxyCRG Episode Task Window.
Operating 
Controls
In the oxyCRG task window you can:
1. Select the previous episode.
2. Select the next episode.
3. Return to the Event Review task window.
4. Record an episode.
5. Print an episode.
6. Delete an event.
Selecting an
Episode
Selecting the 
 and/or 
 softkeys 
jumps from one oxyCRG episode to the next in both directions.
Entering the
Event Review
Task Window
Selecting the 
 softkey returns you to the Event Review 
task window. The event last selected in the oxyCRG Episode task 
window will be selected here.
Previous Episode
Next Episode
Event Review