Adjusting Neonatal Event Review Settings
Neonatal Event Review   9-21
Ne
ona
ta
l Ev
e
n
Revi
ew
  
Desaturation
Event Settings
 Alarm Triggered
The Desaturation events are derived from the first SpO2 bedside module 
and triggered either by a Red*** Desat alarm or a Yellow** low limit SpO2 
alarm. The Desat/SpO2 alarms and trigger times can be viewed but not 
adjusted in the Event Setup task window. These settings must be adjusted 
in Monitoring Mode in the SpO2 task window. If the SpO2 alarm related 
trigger threshold and trigger time items are selected, the 
 
key is inactive and the following status message is 
displayed in the Event Setup task window:
“SpO2 items must be changed in the SpO2 Setup”
User Defined
For situations where the storage and documentation of an event should 
be independent of all Desat/SpO
2
 alarms, you can adjust the Desaturation 
event threshold and trigger time in Monitoring Mode in the Event Setup 
task window.
Desat Trigger Off
It is also possible to turn the Desaturation event triggers OFF in cases 
where the occurrence of Desaturation events are so frequent that the 
event storage facility risks becoming full (a total of 48* events can be 
stored). If the Desaturation event trigger is switched to OFF, the oxygen 
saturation is still displayed in the oxyCRG Episode task window but the 
Desaturation events are not captured or displayed in the Event Review 
task window
*24 for V24 and V26 Patient Monitor
s
Operating 
Controls
In the Event Setup task window you can:
1.  Select items. 
2. Change the item content.
Selecting
Items
Selecting the 
 softkey enables you to step through the items 
listed under each event group.
Change Content
Select Item