Data Transfer Module
10-4     Data Transfer  
D
a
ta
 T
ra
n
sf
er
  
Data transfer presents no interruptions to realtime monitoring and 
provides near-continuous trends. However, some functions under 
 and 
 are not available during the transfer 
of data. The ST Analysis task windows are unavailable during a transfer 
to the monitor.
 The front of the module has two keys:
The 
Monitor
 key for initiating a transfer from the module to the 
monitor. 
The 
Module
 key for initiating a transfer from the monitor to the 
module. The information transferred may consist of up to 24 hours 
of the database, ongoing updates to the database, or both. 
Transfers can also be initiated or stopped through 
.
During a transfer, the module light is on when the corresponding Data 
Transfer task window is active. The light blinks rapidly when a transfer 
is in progress, and at a slower rate during the collection of new data (see 
“Types of Transfer” on page 10-7).
Trends/Calcs
Procedures
M1235A
DATA
TRANSFER
!
MODULE
MONITOR
M1235A
Procedures