Maintenance Checks
Monitor Installation and Patient Safety   11-15
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
This recommendation does not supersede any local procedures or legal 
requirements.
Electrical Safety Tests
At least every two years,  after 
system installation, after any 
repairs, or as needed.
See “Testing and Maintaining the 
CMS” in the “CMS Service Guide” 
or “Testing and Maintaining the 
V24 and V26“in the “V24 and V26 
Service Guide”.
Testing of all cables and cords.
Daily.
See “Testing and Maintaining the 
CMS” in the “CMS Service Guide” 
or “Testing and Maintaining the 
V24 and V26“in the “V24 and V26 
Service Guide”.
Calibration of the measurement 
modules where appropriate
(For EU/EFTA countries only)
Every 2 years.
See individual module chapters in 
Volume 2 of this manual for 
details.
User Tests
Module dependent. Please refer to 
chapter 13 “Maintenance” of this 
manual or chapter 3 of the CMS 
Service Guide “Maintaining the 
CMS” for details.
See chapter 13 “Maintenance” of 
this manual for details.
Maintenance
Frequency
Source of Information