Controls and Connectors
11-22     Monitor Installation and Patient Safety  
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
Note—
Do not apply a voltage of more than 
±
12V to any of these outputs.
The connectors on the rear panel of the computer module (together with 
their connecting function cards) are:
1. System Power Connector; on the M1046A Computer Module, this 
is a connector used to input the 60V dc power supply from the 
CRT display module. It is not for use on the M1046B Computer 
Module but is instead fitted with a stub connector.
2. Equipotential Grounding Terminal; this is used to connect the 
computer module to the hospital´s grounding system.
3. Philips monitoring network connectors (Philips Network 
Interface); these are Philips monitoring network connectors, used 
to input/output information to/from a Philips monitoring network.
4. Human Interface Connector (Utility CPU); this is a Philips HIL 
connector, used to input information from the keypad in the 
display module.
5. Nurse Paging Relay output; this is a mini-phone jack, used to 
connect to the hospital´s Nurse Paging System.
6. Analog Interface Connector; a 37-pin “D” type connector is used to 
output information to analog devices.
7. RS232 Connectors (RS232 Interface); these are 25-pin “D” type 
connectors, used to output information to printers or computers, 
RS232 Connectors
±
12 V
HDLC Connector
5 V
Video Out (9)
5 V
Flatscreen Remote Power Supply On/Off
15 V
Video Out (11)
60 V
Maximum Voltages