Battery Care and Maintenance
Battery Information (V24CT and V26CT only)   12-9
Ba
tte
ry
 I
n
for
m
a
ti
on 
(V
24
CT
 an
d
 V
26C
T
 
Battery Care and Maintenance
The Philips 40488A sealed lead-acid batteries used in the V24CT provide 
optimum performance and battery life when properly cared for. Proper 
handling of batteries is essential to the life of the battery and safety of the 
user.
Storage
Store batteries in a cool dry area. A refrigerator is an excellent place 
to store batteries. Allow batteries to return to room temperature 
before charging or use. 
The charge in the batteries gradually deteriorates even during 
storage. They will self-discharge 4-5% for each month in storage at 
25°C. 
Batteries should be stored in a fully charged state. 
A stored battery loses capacity over time. To minimize loss of 
capacity recharge battery at least once every 6 months. Batteries 
can be charged in the monitor or by using a battery charger. 
Note—
Recharge the batteries after using them. It is especially important 
to avoid allowing the batteries to stay in a fully discharged state for long 
periods of time. Simply plug the monitor into AC line power after 
transporting a patient or when the monitor is not in use. If possible, avoid 
discharging the battery completely. 
Replace batteries with known charged batteries in the following 
conditions:
Batteries discharge quickly when being used
Batteries’ runtime after first “RECHARGE BATTERIES” 
alarm is less than ten minutes
Batteries do not charge to full capacity (indicated by the 
monitor's battery fuel gauge)