Monitor Maintenance
Maintenance   13-9
M
a
int
e
na
nc
e
  
Perform a 
Start-up 
Sequence 
Test of the 
System 
The System automatically starts every time the system is switched ON.
TEST - Turn the System On by pressing the power switch in the lower left-
hand corner of the display.
RESULT - The system will click and light-up the LEDs on the control 
panel. After several seconds, the LEDs will go off and the display will 
start showing waveforms and numerics. If the system fails to boot, notify 
your biomedical department or the Philips representative.
Verify the 
Integrity of 
the Display
Note the position of the contrast control on the V24 (or the brightness 
control on the V24C/V24CT/V26C/V26CT) before starting tests so that they 
are returned to their previous positions (assuming they are correct) at the 
end of the testing procedure.
1. Set up an ECG signal from a fixed rate simulator on the screen. 
2. Check the ECG complexes. They should be regularly spaced. 
Uneven spacing indicates missing pixels. All portions of a simulated 
ECG waveform should be clear and visible, including the P wave 
and QRS. 
3. Adjust the contrast control on the V24 (or the brightness control on 
the V24C/V24CT/V26C/V26CT) so it is at the maximum setting and 
then adjust to the minimum setting. 
4. Check to see if any noise (interference) is superimposed on the 
baseline with the ECG simulator attached. Baseline interference 
may be apparent as a thick baseline at high gain settings. 
5. Apply an external 1 mV pulse. The trace should exhibit a leading 
edge with minimal rounding and a spike (or overshoot) of less than 
10%. After 1 second, the pulse should have decayed no more than 
half of its original amplitude. 
6. Return the contrast control on the V24 (or the brightness control on 
the V24C/V24CT/V26C/V26CT) to its previous position (assuming it 
was correct). If any problems are identified, notify your biomedical 
department or the Philips representative.