Using an ITE Display
The CMS and V24 and V26 Patient Monitors   1-39
EMC
1. The display must either fulfill the requirements of the EN60601-1-2 
{IEC 601-1-2}, or
it must conform to the requirements of the EMC standard for ITE 
devices, EN50082-1: 1997 and EN50081-1 / CISPR 22.
2. The video cable between M1167A/77A CMS Patient Monitoring 
System and the ITE display must not exceed a length of 3.0m. For 
cables exceeding 3.0m EMC testing according to IEC 801-4:1988 has 
to be done.
Perfor-
mance 
Require-
ments
Philips cannot assure compliance with the ANSI/AAMI EC-13 Standard for 
Cardiac Monitors, Heart-rate Meters and Alarms when using ITE displays. 
Compliance with the ECG aspect ration and 25 mm/s 
±
10% sweep-speed 
specifications can only be assured when using the Philips Displays 
M1092A, M1094B and M1095A. In addition, the following display 
specifications are recommended:
Specification
Requirement or Value
Units
Resolution (pixel addressability)
1024 x 768
Dots x lines
Vertical Refresh Rate
60
Hz (non-interlaced)
Red, Green and Blue Video Inputs
~0.7
V p-p
Vertical & Horizontal Multi-Sync Inputs
5
V TTL
Video-Cable Connector
HD15male