CSA Display (CMS only)
Setting up your Monitor   3-29
S
e
tt
ing up y
our
 
M
onit
o
r  
Main Screen
Notes on 
CSA
The CSA information is overlayed on the realtime waves at the 
bottom of the Main Screen. These realtime waves disappear from 
the display, but the parameters are still being monitored, and the 
parameter alarms are still active.
If CSA covers part of an overlap channel, the entire overlap channel 
disappears from the display.
If you are using a touchscreen, selecting the CSA display via touch 
or mouse-click will open up the EEG Setup Task Window
Standard Display
CSA Display
4 realtime waves
CSA + 2 realtime wave
6 realtime waves
CSA + 3 realtime waves
8 realtime waves
CSA + 4 realtime waves