Philips V26CT 用户手册

下载
页码 392
Changing Operating Modes
Setting up your Monitor   3-55
S
e
tt
ing up y
our
 
M
onit
o
r  
Changing Operating Modes
This function will be mainly used by service engineers for configuring the 
system on installation or for servicing the system. Before you can operate 
in this Task Window you have to use a password.
There are three levels of password:
Warning
Warning
The Configuration, Service and Demo modes do not support 
patient monitoring. When the monitor is attached to a patient it 
must be in the monitor mode. 
Level 1 password
 allows you to select Monitor or Config mode only. 
Level 2 password
 allows you to select Monitor or Service mode only. 
Level 3 password
allows you to select any of the four operating 
modes listed below: 
Monitor
This is the usual operating mode and fully supports 
patient monitoring. If the system is attached to a patient 
it must be in this mode. 
Config
 This mode is for configuring the monitor, for example, 
on installation. The settings that are selected in this 
mode will be the active default settings for when the 
system is turned on. 
Service
This mode is used by service personnel. 
Demo
This mode can be used for training purposes. The data 
base has 24 hrs of simulated waveform and trend data.