Using Philips Patient Care System with an Arrhythmia Computer
4-8     Other Patients  
Ot
h
e
r Pa
ti
e
n
ts
  
Using Philips Patient Care System with an Arrhythmia 
Computer
Note—
The patient monitors perform analysis on the patient’s ECG 
waveform. This includes triggering alarms on asystole, bradycardia, 
ventricular fibrillation and tachycardia conditions. Enhanced arrythmia 
monitoring is available when the patient monitor is connected to one of 
Philips’ arrhythmia systems via the Philips monitoring network interface 
(see below).
If your monitor is connected via the Philips patient care system to an 
arrhythmia computer (78720A), or you have a central station with 
arrhythmia channels such as the 78560A or 2300, or you have a M3150A/
M3153A Philips Information Center, you are able to display a delayed 
arrhythmia wave. Using the 78720A or the M3150A/M3153A, the 
arrhythmia monitoring for your monitor must be turned on or off at the 
computer display. Using the 78560A or M2300, this can be done from the 
central station or from the monitor itself.
Arrhythmia monitoring is most easily assigned in SELECT BED at the 
central station or arrhythmia computer, or from the ARRHYTHMIA 
MENU at the monitor. The ARRHYTHMIA MENU enables you to:
Set up arrhythmia monitoring on a patient (not available for 
78720A).
Tabulate trends and review trend data (M2300 only).
Review and edit alarms (not available for 78720A).
When arrhythmia ia assigned, the monitor automatically goes to “Paced” 
mode. You cannot change the mode back to “Non-Paced” mode until 
arrhythmia is unassigned.
Note—
The screens that can be seen on the monitor via the 
ARRHYTHMIA MENU are all similar to screens on the arrhythmia 
computer or central station. For further information on using the 
arrhythmia monitoring function, refer to the operating guide for the 
arrhythmia computer or central station.