Using Philips Patient Care System with an Arrhythmia Computer
4-14     Other Patients  
Ot
h
e
r Pa
ti
e
n
ts
  
** R-ON-T PVCs
For HR <100, a PVC with R-R interval <1/3 
the average interval followed by a 
compensatory pause of 1.25 x average R-R 
interval or 2 such Vs without a 
compensatory pause occurring within 5 min. 
of each other. (When HR >100, 1/3 R-R 
interval is too short for detection.)
** VENT BIGEMINY
A dominant rhythm of N, V, N, V
(N=supraventricular beat, V=ventricular 
beat)
** VENT TRIGEMINY
A dominant rhythm of N, N, V, N, N, V
(N=supraventricular beat, V=ventricular 
beat)
** PVCs > xxx/MIN
PVCs within one minute exceeded the PVCs 
/min limit (xxx)
** MULTIFORM PVCs
The occurrence of two differently shaped 
Vs, each occurring at least twice within the 
last 300 beats as well as each occurring at 
least once within the last 60 beats
** HR yyy > xxx
Heart Rate (yyy) greater than the upper HR 
limit (xxx)
** HR yyy < xxx
Heart Rate (yyy) lower than the lower HR 
limit (xxx)
** IRREGULAR HR
Consistently irregular rhythm (irregular R-R 
intervals)
Message
Minimum Condition Requiring an Alarm