Alarm Display
5-2     Alarm Functions  
Ala
rm Func
tions
  
Alarm Display
The alarm messages are displayed across the top of the screen. When 
you switch the power on, the system alarm capability is switched off. 
This is indicated by an ALARMS SUSPENDED or ALARMS SUSPND 1, 2 
or 3 MIN message at the top right corner of the screen, depending on the 
alarm suspended mode which is configured on the monitor. The Alarm 
Suspend symbol(
) on the control panel is also illuminated.
Note—
When you switch the monitor on, a loudspeaker test is conducted 
for approximately 3 seconds. If there is no sound, please check the 
volume settings.
If there is more than one red alarm, yellow alarm or INOP message, you 
will see an up arrow (
) next to the message. The messages are 
displayed in rotation every 2 seconds. If your patient has more than one 
alarm, you can see a list of all these active alarms in the Alarm Messages 
Task Window. To access this Task Window press 
 or 
 followed by 
.
Warning
Warning
Alarms and INOPs are not active in the configuration and service 
modes, but are active in monitoring and demo modes. 
Alarm 
Functions 
on the 
Control 
Panel
Alarms/Volume
 
Alarms
 
Alarm Messages
Alarms/
Other
Monitor
Setup
Main
Screen
Change
Screen
Silence/
Reset
Alarm Lamps
Alarms Suspended Lamp
Suspend
Silence/Reset key
Suspend key
Volume