Alarm Display
5-10     Alarm Functions  
Ala
rm Func
tions
  
Individual 
Parameter 
Alarms
When you plug in a parameter module and connect the transducer or 
adapter cable, the parameter is automatically switched on (ECG, RESP 
and NBP are automatically switched on when you plug in the relevant 
module).
Whether the parameter alarms are switched on or off at start-up can be 
selected for each parameter separately in the Configuration Mode.
The parameter alarms can be switched off individually either in each 
parameter Adjust Alarms Task Window or in the Alarm Limits Task 
Window. A 
 symbol appears next to the numeric of any parameter for 
which the alarms are switched off. A parameter without any INOP or 
alarm capabilities does not show a 
If all the alarms have been turned off individually, they must be turned 
on individually. There is one exception to this rule, when the monitor is 
set up in Operating Room mode (this applies to all ACMSs and V24 and 
V26 Patient Monitors and others using a Configuration Set with Unit 
Type=OR). In this mode all parameter alarms can be switched off with a 
single key press. This enables you, for example, to quickly switch all 
alarms off and individually switch on those parameter alarms really 
needed, depending on your monitoring requirements. All parameter 
alarms can be switched off (or on) using 
 in the Alarm Limits 
Task Window.
Note—
We recommend not to switch off all parameter alarms at once and 
then switch on individual alarms, but rather to switch off individual 
parameter alarms, applying extra vigilance to these parameters.
If a parameter alarm is turned off, INOP messages continue to be 
displayed but alarm messages, lamps and the alarm tones are not active. 
The nurse call relay is inactive.
Note—
The heart rate derived from a continuously monitored ECG is a 
measure of the heart's overall electrical activity. The pulse rate is a 
measure of the heart's mechanical activity, pumping blood around the 
body. With the HR/PULSE parameter you can select only one of them to 
be the active alarm source. If you choose HR, all alarms for PULSE are 
not active. If you choose PULSE, all alarms for HR are not active. The 
INOPs for both HR and PULSE are displayed regardless of which one is 
the alarm source. 