Viewing Patient Data
Trends and Calculations   8-3
Tr
e
nds
 a
nd 
C
a
lcula
ti
o
n
s
  
Viewing Patient Data
There are two ways to view patient data:
In either form, the most recent data is displayed at the right of the screen 
(unless entered through the Graph Trends where you may have moved 
the time bar). The patient data will be presented in the default resolution 
(level of detail) and time span.
The number of parameters available to be viewed depends on the patient 
data management configuration and the options purchased. Data is stored 
and appears in tables and graphs (if configured) even though the numeric 
is not displayed on the Main Screen. 
The number of parameters tracked, the frequency of measurement 
capture, and the number of hours of database storage are determined in 
configuration mode prior to monitor use. The combinations are:
Standard Database 
In graphs
where parameter values are points plotted on a graph 
along a time axis
In tables
where parameter values and the times the values were 
collected are listed in a table 
No. of Parameters 
Resolution 
Size 
16 
1 minute   
24 hours
16 
12 seconds 
 4 hours