Installation Information
Monitor Installation and Patient Safety   11-9
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
Environ-
ment
To ensure a completely safe electrical installation, follow the instructions 
described later in this section. The environment where the system will be 
used should be reasonably free from vibration, dust, corrosive or 
explosive gases, extremes of temperature, humidity, and so on.
For a cabinet mounted installation, allow sufficient room at the front for 
operation and sufficient room at the rear for servicing with the cabinet 
access door open.
Approximately 15 minutes after switch-on, the patient monitor and the 
Anesthetic Gas Module operate within specifications at ambient 
temperatures shown in the tables below. These specifications do not 
apply to the plug-in modules; they are specified for different ambient 
temperatures.
Ambient temperatures that exceed these limits could affect the accuracy 
of these instruments and cause damage to the components and circuits. 
Allow at least 2 inches (5cm) clearance around the instruments for proper 
air circulation.
Refer to the tables below for a summary of the environmental 
specifications of the different system combinations: