Maintenance Checks
11-16     Monitor Installation and Patient Safety  
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
Patient 
Cables and 
Leads
Inspect the patient cables and leads and their strain reliefs for general 
condition. Examine cables carefully to detect breaks in the insulation 
and to ensure that they are gripped securely in the connectors of each 
end to prevent rotation or other strain.
To ensure operation of the Anesthetic Gas Module within specified 
limits:
calibrate the Anesthetic Gas Module at least once every 12 
months, or whenever the validity of the readings is in doubt.
check the ventilator fan in the Anesthetic Gas Module for proper 
operation every 6 months.
All checks which require the instrument to be opened must be made by 
qualified hospital service personnel. Safety and maintenance checks can 
also be made by Philips personnel.
Warning
Warning
Failure on the part of the responsible individual hospital or 
institution employing the use of this equipment to implement a 
satisfactory maintenance schedule may cause undue equipment 
failure and possible health hazards.