Philips V26CT 用户手册

下载
页码 392
General Disinfecting of the System
13-4     Maintenance  
M
a
int
e
na
nc
e
  
General Disinfecting of the System
We recommend that you disinfect the product only when necessary as 
determined by your hospital’s policy, to avoid long term damage to the 
product. 
We also recommend that the products being disinfected be cleaned first, 
as described under “General cleaning of the System” on page 13-2. 
Use the recommended disinfecting substances listed below:. 
Caution
To avoid damage to the product, observe the following general 
precautions for disinfection. You should only deviate when this is 
explicitly stated in the disinfecting instruction of a specific product.
1. Do NOT use Povodine
®
, Sagrotan
®
, Mucocit
®
 disinfecting agents 
or strong solvents e.g. acetone.
2. Always dilute according to the manufacturer’s instructions or use 
lowest possible concentration.
3. Do not allow any liquid to enter the case. 
4. Never submerge any part of the system.
Alcohol 
based 
Ethanol 70%, Isopropanol 70%
 
Aldehyde 
based
Cidex
Bleach
Dilution of sodium hypochlorite (laundry bleach): concentration 
ranging from 500ppm (1:100 dilution of household bleach) to 
5000ppm (1:10 dilution of household bleach), 
Hydrogen peroxide 3%,  Chlorox
 (1:10 dilution), 
Dakin’s Solution.
Phenol based
Wofasept
, Sporicidin
.