Performance Assurance Checks
13-20 Maintenance
M
a
int
e
na
nc
e
Performing
the Blood
Analysis
Module Self-
Test
the Blood
Analysis
Module Self-
Test
Auto Check
- The Blood Analysis Module performs a check when the
system is switched on
Self-Test
- The Self-Test with the External Simulator M3634A may be
performed at any time.
TEST - Insert the External Simulator into the cartridge slot of the
module.
module.
RESULT - Either PASS or FAIL will be displayed on the Monitor to
indicate whether the module has passed the self-test or not. If the
module fails the self-test, try the procedure again. If it fails again, wait
half an hour and then try the procedure again. If the module continues
to fail the self-test the module should be replaced.
indicate whether the module has passed the self-test or not. If the
module fails the self-test, try the procedure again. If it fails again, wait
half an hour and then try the procedure again. If the module continues
to fail the self-test the module should be replaced.
Performing
the
Recorder
Module Self-
Test
the
Recorder
Module Self-
Test
Auto Check
- The recorder module performs a check when it is plugged
into the parameter module rack.
Self-Test
- The Recorder Module check may be performed at any time.
TEST - Turn on the system power and plug in the recorder module
RESULT - The LED on the front panel indicates whether the module has
passed the self-test by flashing.
passed the self-test by flashing.
LED flashes twice - module has passed check LED flashes once - module
failed self-test. Take the module out and try the procedure again to
make sure there is a good connection. LED does not flash - module
processors are not working properly and the module should be
replaced
failed self-test. Take the module out and try the procedure again to
make sure there is a good connection. LED does not flash - module
processors are not working properly and the module should be
replaced
Performing
the Data
Manage-
ment Data-
base Self-
Test
the Data
Manage-
ment Data-
base Self-
Test
The data management database is tested during system start-up with an
SRAM checksum procedure. If there is an error found in the SRAM
checksum, the system will restart (cold start) and all patient data will be
lost.
SRAM checksum procedure. If there is an error found in the SRAM
checksum, the system will restart (cold start) and all patient data will be
lost.