Performance Assurance Checks
13-20     Maintenance  
M
a
int
e
na
nc
e
  
Performing 
the Blood 
Analysis 
Module Self-
Test
Auto Check
 - The Blood Analysis Module performs a check when the 
system is switched on
Self-Test
 - The Self-Test with the External Simulator M3634A may be 
performed at any time.
TEST - Insert the External Simulator into the cartridge slot of the 
module.
RESULT - Either PASS or FAIL will be displayed on the Monitor to 
indicate whether the module has passed the self-test or not. If the 
module fails the self-test, try the procedure again. If it fails again, wait 
half an hour and then try the procedure again. If the module continues 
to fail the self-test the module should be replaced.
Performing 
the 
Recorder 
Module Self-
Test
Auto Check
 - The recorder module performs a check when it is plugged 
into the parameter module rack. 
Self-Test
 - The Recorder Module check may be performed at any time.
TEST - Turn on the system power and plug in the recorder module
RESULT - The LED on the front panel indicates whether the module has 
passed the self-test by flashing.
LED flashes twice - module has passed check LED flashes once - module 
failed self-test. Take the module out and try the procedure again to   
make sure there is a good connection. LED does not flash - module 
processors are not working properly and the module should be   
replaced
Performing 
the Data 
Manage-
ment Data-
base Self-
Test
The data management database is tested during system start-up with an 
SRAM checksum procedure. If there is an error found in the SRAM 
checksum, the system will restart (cold start) and all patient data will be 
lost.