Indications for Use
    
      ix
Exposure of electrical contacts or connections to saline or other liquids and gels is 
dangerous. Electrical contacts and connections such as cable connectors, power 
supplies, parameter module plug-in connections and rack connections must be kept 
clean and dry. Thoroughly dry any electrical connections that become contaminated 
with liquids. If additional decontamination is required please contact your biomedical 
department or Philips Medical Systems Response Center. 
Although this equipment is shielded against Electromagnetic Interference (EMI), it is 
recommended to avoid the use of electrically radiating devices in close proximity to this 
equipment. 
Connecting the Philips monitoring network (SDN) cable when the product is powered on 
is not supported. Error codes and Philips monitoring network (SDN) interface lock-up 
may occur. Power cycling the product will recover the product. No permanent damage 
will result. To prevent unintentional disruption in monitoring, be sure the SDN interface 
cable is properly secured at both ends when connecting to the Philips monitoring 
network (SDN).
Do not connect a second rack by a cable when using a module rack docked to the back 
of the V24CT or V26CT. Using a second rack connected by a cable may disrupt module 
communication.
Caution
A “caution” calls attention to a condition or possible situation that could cause injury to the 
user. 
Ventilation Requirements - Failure to meet ventilation requirements may cause 
equipment failure and, in turn, jeopardize the functions of automated monitoring. Do not 
locate equipment in an enclosed area which could restrict heat dissipation.
Maintenance - Failure on the part of the responsible individual, hospital, or institution 
employing the use of this equipment to implement a satisfactory maintenance schedule 
may cause undue equipment failure and possible health hazards.
Do not spray cleaning solutions directly onto the patient monitor. Moisture droplets may 
enter the internal components and cause equipment malfunction or failure. Cleaning 
solutions should be applied to a cloth and the cloth used to wipe the monitor clean. The 
monitor should be turned off during cleaning.