Chapter 3
Signal Connections
© National Instruments Corporation
3-21
Figure 3-11.  Pretrigger DAQ Timing
For interval scanning data acquisition, counter B1 determines the scan 
interval. Instead of using counter B1, you can externally time the scan 
interval through OUTB1. If you externally time the sample interval, we 
recommend that you also externally time the scan interval.
Figure 3-12 shows an example of an interval scanning DAQ operation. The 
scan interval and the sample interval are being timed externally through 
OUTB1 and EXTCONV*. Channels 1 and 0 of the input multiplexers are 
scanned once during each scan interval. The first rising edge of 
EXTCONV* must occur a minimum of 50 ns after the rising edge on 
OUTB1. The first rising edge of EXTCONV* after the rising edge of 
OUTB1 enables an internal GATE signal that allows conversions to occur. 
The first conversion then occurs on the following falling edge of 
EXTCONV*. The GATE signal disables conversions for the rest of the scan 
interval after the desired channels have been scanned. Refer to th
 section in Chapter 4, 
, for 
more information on interval scanning.
t
w
  50 ns minimum
EXTTRIG
EXTCONV*
CONVERT
V
IH
V
IL
t
w
t
w