National Instruments 1142 用户手册

下载
页码 86
Chapter 5
Using the SCXI-1141/1142/1143 Module
5-6
ni.com
Table 5-3.  NI-DAQmx RTD Measurement Properties
Property
Short Name
Description
Analog Input»Temperature» 
RTD»Type
AI.RTD.Type
Specifies the type of RTD 
connected to the channel.
Analog Input»Temperature» 
RTD»R0
AI.RTD.R0
Specifies the resistance in ohms of 
the sensor at 0 
°C.
Analog Input»Temperature» 
RTD»Custom»A, B, C
AI.RTD.A
AI.RTD.B
AI.RTD.C
Specifies the A, B, or C constant of 
the Callendar-Van Dusen equation 
when using a custom RTD type.
Analog Input»General Properties» 
Signal Conditioning»Resistance 
Configuration
AI.Resistance.Cfg
Specifies the resistance 
configuration for the channel, such 
as 2-wire, 3-wire, or 4-wire.
Table 5-4.  NI-DAQmx Thermistor Measurement Properties
Property
Short Name
Description
Analog Input»Temperature» 
Thermistor»R1
AI.Thrmistr.R1
Specifies the resistance in ohms of 
the sensor at 0 
°C.
Analog Input»Temperature» 
Thermistor»Custom»A, B, C
AI.Thrmistr.A
AI.Thrmistr.B
AI.Thrmistr.C
Specifies the A, B, or C constant 
of the Steinhart-Hart thermistor 
equation, which NI-DAQmx uses to 
scale thermistors.
Table 5-5.  NI-DAQmx Current Measurement Properties
Property
Short Name
Description
Analog Input»General Properties» 
Signal Conditioning»Current 
Shunt Resistors»Location
AI.CurrentShunt.Loc
Specifies whether the 
shunt resistance location is 
internal or external.
Analog Input»General Properties» 
Signal Conditioning»Current 
Shunt Resistor»Value
AI.CurrentShunt.Resistance
Specifies the resistance, in 
ohms, of the external shunt 
resistance.